Abstract

In personnel monitoring services, it is important to omit the high-temperature annealing process so that large numbers of TL detectors can be produced economically. There are two efficient ways of reducing the residual signal of LiF:Mg,Cu,P. One is by increasing the maximum readout temperature and the other is by improving the preparation procedure (increasing the Cu concentration and the sintering temperature) but both reduce the TL sensitivity. In personal dosimetry the real dosimetric signals are separated from the residual signals by computerised analysis of glow curves. The adverse influence of the high residual signals of LiF:Mg.Cu.P TL material has been effectively eliminated and the sensitivity remains stable. A good dosimetric result using only reader measurement without pre-irradiation oven annealing is attained in a dose range of 50-80,000 microGy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.