Abstract
The static characteristics of platinum resistance thermometers with different platinum purities were measured. It is shown that in the 0–420°C range the ITS-90 method gives an error of less than 0.01°C, while in the 0–230°C range the error is less than 0.006°C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.