Abstract

High-energy accelerators can produce strong time-structured radiation fields. Such dose shots are generated at linear machines with low duty cycles as well as at circular machines when complete fills are instantaneously lost. The main dose component behind thick shielding is due to high-energy neutrons occurring at that time structure. Dosemeters based on Geiger-Mueller tubes or proportional counters fail here completely. The 12B counter, a novel dosemeter made of a plastic scintillator using carbon activation for event-like exposure, has been introduced. High-energy neutrons activate the carbon nuclei by three inelastic reactions. The decay patterns with half-lives between 20 ms and 20 min can be exploited depending on the time structure of the radiation field. The response of the 12B counter was measured along with some other dosemeters, both active and passive, in the radiation field behind the lateral concrete shielding of a 7.5 GeV proton transfer line.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.