Abstract

This article deals with the development and comparison of eight different electrodes made out of a cotton fabric substrate, a silver coated yarn and partly conductive finishes, i.e. a PEDOT:PSS Orgacon ICP 1050 dip-coating and a Powersil coating. The purpose is the application especially in the medical field of angiopathy like for bioimpedance measurements during compression therapies. To be able to compare the suitability of the electrodes, various tests have been performed of the coating abrasion resistance, the stability of electrical resistance values, as well as resistance and bioimpedance measurements. Significant differences between the electrodes regarding their resilience and resistance that are visualized in a value-added analysis were found, with one hand-embroidered, one machine-sewn and one commercial electrode showing optimum properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.