Abstract

The micro-EDXRF (energy dispersive X-ray fluorescence) method was applied to the screening of Pb in micrometer-area samples, such as a Cu contact in electrical components that had been coated by Pb-free Sn-Ag-Cu solder. The reliability of the screening method was evaluated by a comparison with a scanning electron microscope (SEM) observation and a precious chemical analysis method of inductively coupled plasma mass spectrometry (ICP-MS). Some factors that affect the testing reliability, such as the thickness of the solder, the segregation of Pb and Ag, etc. were found by SEM observations. By adjusting some calculation parameters, screening of the micrometer area (0.1 mm) was performed using the fundamental parameter (FP) method for a thin film in conjunction with micro-EDXRF. The measurement error ranged by 25% for the thin film-FP method. The resulted detection limit was 0.04 wt% for Pb, depending on the solder thickness. This method can be successively applied for quality control to check the purity of a Pb-free Sn-Ag-Cu solder coating in electrical components.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.