Abstract

We assessed the applicability of tendon excursion measurement by means of Colour Doppler Imaging (CDI) on human specimens, and also assessed the correlation between values measured by Doppler and by displacement meters. Muscles were separately connected to a mass of 1 kg with a steel wire running over a pulley. This weight moved the telescopic end of a digital displacement meter up and down during passive extension and flexion of the fingers. Excursion was measured with a pulsed multi-channel CDI scanner on the same arm. Assessment of finger tendon excursion with CDI correlated well with the mechanical micro displacement meter, the latter being considered the most accurate method in cadaver studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.