Abstract

ABSTRACTInfrared thermometry is a widely used technique for contactless temperature measurement, which is often conducted through semi-transparent media. In the present work, influences on the measurement results stemming from semi-transparent media and from the optical characteristics of the measurement setup are discussed. Results of two experimental setups, containing low, medium, and high transmission media are presented and compared to calculated data using a one-dimensional analytical approach and a three-dimensional ray-tracing algorithm. It is shown through modeling and experiments that the surroundings and, in particular, the (semi)transparent materials within the optical path are critical for accurate temperature measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.