Systematic patient-dose errors relating to standard equivalent-square approximations and movable collimator-exchange effect were evaluated for rectangular fields of photon beams from 4- and 10-MeV microwave electron accelerators (Varian Clinac 4 and Clinac 18). Errors approached 2.5% at nominal treatment depths for highly elongated rectangular fields. The collimator-exchange effect should be considered when (a) evaluating ionization-chamber stem effect, (b) specifying relative output factors for rectangular fields, and (c) evaluating equivalent-square-field algorithms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.