Abstract

The fatigue strength of three peripheral suture techniques for flexor tendon repair was compared by cyclic loading of the repairs in the porcine flexor digitorum tendon. Thirty-six tendons were sutured using only peripheral sutures with 6-0 Nylon. An initial cyclic load of 10 N for 500 cycles was applied and increased by 10 N for an additional 500 cycles at each new load until rupture. The fatigue strength of the symmetric running peripheral suture was 85.0% and 144.8% greater than that of the two kinds of the asymmetric running peripheral sutures. Symmetric running sutures can enhance the suture strength and appears to be a useful technique for increasing the strength of the peripheral suture.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.