Abstract
To compare thin-section, sagittal, fast fluid-attenuated inversion-recovery (FLAIR) magnetic resonance (MR) imaging with conventional axial spin-echo (SE) imaging for early detection of multiple sclerosis (MS) in the brain. Conventional 5-mm axial proton-density- and T2-weighted SE images and sagittal 2-mm fast FLAIR images of the brain in five healthy volunteers and 25 patients with clinically suspected MS were evaluated and graded as normal or as showing possible or probable MS in a double-blind study. Fast FLAIR was judged better than SE in 37% of cases, equal to SE in 63%, and never worse than SE. In 43% of patients considered to have normal SE images, fast FLAIR images showed abnormalities consistent with MS. A subependymal striated appearance observed in several cases is believed to represent early inflammation and/or demyelination around subependymal veins in MS. Sagittal thin-section fast FLAIR is superior to conventional axial proton-density- and T2-weighted SE pulse sequences for detection of MS plaques in the brain.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.