Abstract

Superior vena cava (SVC) isolation during atrial fibrillation catheter ablation is limited by the risk of collateral damage to the sinus node and/or the phrenic nerve. Due to its tissue-specificity, we hypothesized the feasibility and safety of pulsed-field ablation (PFA)-based SVC isolation. One hundred and five consecutive patients undergoing PFA-based AF catheter ablation were prospectively included. After pulmonary vein isolation (±posterior wall isolation and electrical cardioversion), SVC isolation was performed using a standardized workflow. Acute SVC isolation was achieved in 105/105 (100%) patients after 6 ± 1 applications. Transient phrenic nerve stunning occurred in 67/105 (64%) patients but without phrenic nerve palsy at the end of the procedure and at hospital discharge. Transient high-degree sinus node dysfunction occurred in 5/105 (4.7%) patients, with no recurrence at the end of the procedure and until discharge. At the 3-month follow-up visit, no complication occurred. SVC isolation using a pentaspline PFA catheter is feasible and safe.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.