Abstract
The 200 ms paced AV delay is equal to the native AV conduction time, which leads to the phenomenon called ventricular pseudofusion as the ‘paced QRS’ morphology is identical to the native QRS morphology (QRS complexes marked with an asterisk, Fig. 1) [1]. This is an important observation for two reasons. First, the pacemaker should be reprogrammed to an extended paced AV delay to avoid unnecessary ventricular pacing. Secondly, the presence of ventricular pseudofusion is the clue to answer the question about the short paced AV delay of the third beat. Fig. 1 Postoperative standard 12-lead ECG (1 mV/10 mm, 25 mm/s paper speed). The pseudofusion QRS complexes are marked with an asterisk. See text for details There are various pacemaker features that can shorten the paced AV delay from the set parameter, e.g. ventricular safety pacing [2]. However, in this case, there is no need for knowledge about pacemaker features. If this was a normal pacemaker feature, we should see fully paced QRS morphology when the pacemaker shortens the AV delay to 110 ms (third beat on Fig. 1). The only potential physiological explanation is a junctional ectopy, occurring exactly on time with the ventricular pacing spike, exactly at the time the paced AV delay is shortened, which is extremely unlikely. Therefore, this has to be an artifact. When one examines the printed text above the ECG, the letters are squeezed together right above the part of the tracing with the shortened paced AV delay (rectangular frame in Fig. 1). The ECG machine intermittently stopped rolling the paper for about 90 ms at a time, effectively cutting out 90 ms segments from the ECG tracing. Since the paper had a pre-printed grid, this unusual artifact was not easily detectable.
Highlights
The ventricular and atrial passive fixation pacemaker leads were positioned in the right ventricular apex and right atrial appendage, respectively
The sensing and capture threshold values were satisfactory in both chambers
The device was programmed to DDDR mode 60–130 ppm with both the upper tracking rate and maximum sensor rate at 130 ppm, paced AV delay at 200 ms, and sensed AV delay at 180 ms
Summary
Sudden shortening of the paced AV delay: is this normal pacemaker function? A 74-year-old female patient underwent a Medtronic KDR601 dual chamber pacemaker implantation for syncope due to sinus node disease.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.