Abstract

There are several options for LHB tenodesis; yet, there is no standard of care. This technical note describes an extramedullary all-suture anchor technique for LHB tenodesis that is similar to the extramedullary cortical button technique. The LHB tenodesis is performed by using the Y-Knot (1.3-mm; ConMed Linvatec, Largo, FL).The biceps tenotomy is completed arthroscopically, and a standard subpectoral approach is used for the tenodesis. A reamer is first used to drill through the anterior cortex of the humerus; subsequently, a 1.3-mm drill bit is used to drill through the posterior cortex. The Y-Knot anchor is passed through the bone tunnel and secured on the posterior cortical bone. A modified rolling hitch suture is placed 10 mm distal to the end of the LHB tendon by using one suture limb of the Y-Knot anchor. The other suture limb is pulled to shuttle the LHB tendon into the humerus, and the construct is fixed by tying down one limb to the other. This technical note describes an alternative method for subpectoral biceps tenodesis and uses a small drill hole, conserves bone, and minimizes trauma to the tendon. V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.