Abstract

This paper presents a minimally invasive needle-shaped probe capable of monitoring the electrical impedance of living tissues. This microprobe consists of a 160μm thick SU-8 substrate containing four planar platinum (Pt) microelectrodes. We design the probe to minimize damage to the surrounding tissue and to be stiff enough to be inserted in living tissues. The proposed batch fabrication process is low cost and low time consuming. The microelectrodes obtained with this process are strongly adhered to the SU-8 substrate and their impedance does not depend on frequency variation. In vitro experiments are compared with previously developed Si and SiC based microprobes and results suggest that it is preferable to use the SU-8 based microprobes due to their flexibility and low cost. The microprobe is assembled on a flexible printed circuit FPC with a conductive glue, packaged with epoxy and wired to the external instrumentation. This flexible probe is inserted into a rat kidney without fracturing and succeeds in demonstrating the ischemia monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.