Abstract

This paper investigates a method for measuring SO2 concentration using Fabry-Perot interferometer correlation spectroscopy. In this method, the experimental system is designed as a separated beam, with the beam entering the F-P cavity at two incidence angles simultaneously to match the peak and valley positions of the SO2 absorption cross-section. The system achieves a 2σ detection limit of 28.2 ppm·m(15 cm) at a sampling frequency of 10 Hz. An outfield comparison experiment with the differential optical absorption spectroscopy method shows good agreement for the simultaneous measurement of SO2 concentration from sulfur combustion, with a correlation coefficient of R2 = 0.93. This study introduces a non-dispersive, highly accurate, and fast gas detection technique.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.