Abstract
In this work we tried to clarify the relation between the signal detected by indirect UV mode and the concentration of the analyte introduced by electrokinetic (EK) injection from samples having small but different conductivities. A simple external calibration procedure was proposed for quantitative determination of traces that are present in low-conductivity samples. It was found that measuring the current through the capillary filled with the samples provides useful correction factor for the external calibration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.