Abstract

Rocking curves of micro-channel (MC) water-cooled monochromators are broadened by stresses introduced during fabrication and under X-ray thermal load. This is a problem which will be even more serious with the rise of the fourth-generation synchrotron sources, i.e. the free-electron lasers. The X-ray optics group at the Institute of Physics at the ASCR v.v.i. in Prague is designing, testing and, with company Polovodice a.s., fabricating novel internally water-cooled Si monochromators. Here three new micro-channel geometries are introduced which reduce rocking-curve enlargement owing to the fabrication to less than 2.5 microrad ( approximately 0.5 arcsec). All three MC designs show less rocking-curve enlargement and smoother topographic images. The designs also show better cooling efficiencies than the classical MC design in finite-element analysis calculations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.