Abstract
Matrix effect (ME) is always a major issue for the development of LC–MS/MS method. ME resulting from co-eluting residual matrix components can affect the ionization efficiency of target analytes, leading to quantification errors of the analytes of interest. The present work evaluates MEs of milk samples on simultaneous analysis of four aminoglycosides residues via LC–ESI/MS/MS including streptomycin, dihydrostreptomycin, spectinomycin and kanamycin. Approaches to reduce MEs were examined: optimization of the sample preparation, sample dilution and lower flow rate used. Three commercial sorbents were tested including Oasis MCX, Oasis HLB and Oasis WCX. WCX behaved better for all analytes, but high MEs (80.8–134.9%) were obtained. Therefore, a consecutive SPE of tC18-WCX was found to effectively reduce ME. Milk samples from different manufacturers were analyzed and low MEs (85.6–112.9%) were obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.