Abstract

Screening methods capable of identifying DNA samples that will not yield short tandem repeat (STR) profiles are desired. In the past, quantitation methods have not been sensitive enough for this purpose. In this study, low level DNA samples were used to assess whether Quantifiler has a minimum quantitation value below which STR profiles would consistently fail to be detected. Buccal swabs were obtained and the DNA extracted, quantified, and serially diluted to concentrations ranging from 0.002 to 0.250 ng/microL. Samples were analyzed once with Quantifiler, followed by Profiler Plus amplification and capillary electrophoresis analysis. An absolute minimum value below which STR results were unobtainable could not be defined. From the 96 low level samples tested, STR loci (including one full profile) were successfully amplified and detected from 27% of the samples "undetected" by Quantifiler. However, no STR alleles were detected in 73% of these "undetected" samples, indicating that Quantifiler data may be useful for predicting STR typing success.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.