Abstract

AbstractStirring rates in heterogeneous catalytic reactions have an effect on reaction rates. When conducting small‐scale surveys using a single central stir plate, reaction vessels in different positions experience slightly different levels and patterns of agitation. We probed this effect by running the same reaction 40 times, varying the stir rate (fast/slow) and the vial position using two 3D‐printed vial holders. We found variability of conversion (measured mass spectrometrically) to be approximately two times higher for vials placed somewhat randomly in comparison to when vials are consistently placed at a fixed distance from the centre of the stir plate, but the effect was relatively small and could be minimized by using a high stir rate. For those experimenters wishing to completely eliminate differential stirring as a cause for variation in results, the 3D‐printed circular array we designed is recommended over a conventional rectangular array.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.