Abstract

The NMISA participated in the international key comparison of the pure beta-emitter Technetium-99, CCRI(II)-K2.Tc-99. The comparison solution was standardized using three methods, namely the TDCR efficiency calculation method, the CIEMAT/NIST efficiency tracing method and the 4π(LS)β–γ coincidence tracing method with Co-60 as tracer. Excellent agreement between results obtained with the three methods confirmed the applicability of the beta spectral shape given by the latest (2011) DDEP evaluation of Tc-99 decay data, rather than the earlier (2004) evaluation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.