Abstract
An improved bioassay for fluconazole was developed. This assay is sensitive in the clinically relevant range (2 to 40 micrograms/ml) and analyzes plasma, serum, and cerebrospinal fluid specimens; bioassay results correlate with results obtained by high-pressure liquid chromatography (HPLC). Bioassay and HPLC analyses of spiked plasma, serum, and cerebrospinal fluid samples (run as unknowns) gave good agreement with expected values. Analysis of specimens from patients gave equivalent results by both HPLC and bioassay. HPLC had a lower within-run coefficient of variation (less than 2.5% for HPLC versus less than 11% for bioassay) and a lower between-run coefficient of variation (less than 5% versus less than 12% for bioassay) and was more sensitive (lower limit of detection, 0.1 micrograms/ml [versus 2 micrograms/ml for bioassay]). The bioassay is, however, sufficiently accurate and sensitive for clinical specimens, and its relative simplicity, low sample volume requirement, and low equipment cost should make it the technique of choice for analysis of routine clinical specimens.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.