Abstract

Improvements have been made to a high-stability quartz-crystal microbalance for use in a typical surface-science, ultrahigh vacuum chamber, with a frequency stability of one part in 10(8) (0.06 Hz) over 0.5 h. This gives a resolution equivalent to 2% of an atomic monolayer of oxygen over 0.5 h. The quartz-crystal microbalance (QCM) crystal can be rotated to different surface-analysis positions in the chamber. These characteristics open up the combination of surface and bulk adsorption studies on the same sample without transferring the sample to another chamber. To accomplish this, it was necessary to stabilize the sample temperature to ±0.03 K over several hours. The oscillator performance is illustrated by the uptake of oxygen by a gold-plated QCM crystal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.