Abstract

The flexibility of mobile electron accelerators, which are designed to be transported to an operating room and plugged into a normal 3-phase outlet, make them ideal for use in intraoperative radiation therapy. However, their transportability may cause trepidation among potential users, who may question the stability of such an accelerator over a period of use. In order to address this issue, we have studied the short-term stability of the Mobetron system over 20 daily quality assurance trials. Variations in output generally varied within +/-2% for the four energies produced by the unit (4, 6, 9, and 12 MeV) and changes in energy produced an equivalent shift of less than 1 mm on the depth-dose curve. Hours of inactivity, with the Mobetron powered on for use either throughout the day or overnight, led to variations in output of about 1%. Finally, we have tested the long-term stability of the absolute dose output of the Mobetron, which showed a change of about 1% per year.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.