Abstract

An isocratic stability indicating reversed-phase liquid chromatographic determination was developed for the quantitative determination of lacosamide in the pharmaceutical dosage form. A Hypersil C-18, 4.5?m column with mobile phase containing acetonitrile-water (20:80, v/v) was used. The flow rate was 1.0 mL min-1 and effluents were monitored at 258 nm. The retention time of lacosamide was 8.9 min. The method was found to be linear in the concentration range of 5-100 ?g/ml and the recovery was found to be in the range of 99.15 - 100.09 %. The limit of detection and limit of quantification were found to be 2 ?g/ml and 5 ?g/ml, respectively. Lacosamide stock solutions were subjected to acid and alkali hydrolysis, chemical oxidation and dry heat degradation. The drug was found to be stable to the dry heat and acidic condition attempted. The proposed method was validated and successfully applied to the estimation of lacosamide in tablet dosage forms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.