Abstract

Hyperpolarized (HP) 129Xe NMR and MRI have enabled 129Xe studies with extraordinarily enhanced sensitivity, stimulating new developments in magnetic resonance in chemistry, physics, biology and medicine. However, the standard method of HP 129Xe production inevitably demands Rb vapor for the excitation, which has made the method very sensitive to impurities such as water or oxygen. This is the case especially in the recirculating system. In the present study, stability of the hyperpolarizing system is discussed by proposing the "cell decay constant", which symbolizes the decay rate of the NMR signal obtained from the system. The cell decay constant is effectively decreased to 1/3 by introducing separated chambers and mechanical stirring of the alkali metals used in the system, making it effective for accumulating FIDs over 30 to 100 h. The newly developed hyperpolarizing system has been successfully applied for newly detecting a broad signal at 190 ppm with an industrial material Nanofiber.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.