Abstract

Recently we introduced Top Orthogonal to Bottom Electrode (TOBE) Capacitive Micromachined Ultrasound Transducers (CMUTs) as novel 2D arrays capable of 3D ultrasound imaging using only N transmit/receive and N bias channels for an NxN array. Here we adapt these devices for photoacoustic imaging by biasing one column and receiving across rows in parallel, then biasing another column and receiving along rows, and repeating this procedure until all columns have been biased in turn. Only elements in rows which are biased are dominantly sensitive to receive signals. Additionally, we propose a method to bias more than one column at a time using a biasing pattern sampled from rows or columns of a Hadamard S-matrix to improve signal-to-noise ratio, and demonstrate simulation results validating the proposed theory.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.