Abstract
Super-paramagnetic iron oxide-enhanced magnetic resonance imaging (SPIO-MRI) is highly sensitive for liver tumors. This trial examined the optimal parameters of the fast recovery fast spin echo (FRFSE) method in SPIO-MRI. The FRFSE method is a pulse sequence with recovery of the compulsory longitudinal magnetization. The FRFSE method used in SPIO-MRI has better sensitivity for liver tumor, better tumor contrast, and can be used with respiratory gating. The phantoms used included a ferucarbotran phantom, water phantom, and gelatin phantom. Each phantom was scanned by the FRFSE method and fast spin echo (FSE) method, with changing of scanning parameters, and each CNR was measured. Each CNR was given a point, and points were averaged according to each parameter. Parameters that obtained high average points were considered optimal parameters. The optimal parameters were set to TE=50 ms in the FRFSE method. The scan parameters optimized from the phantom study were used in a clinical case, and good results were achieved, as in the phantom study. We optimized the scan parameters of the FRFSE method and were able to suggest the usefulness of the FRFSE method in SPIO-MRI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.