Abstract

Spectral acquisition time is one of the limiting factors in electron paramagnetic resonance (EPR) retrospective biodosimetry in teeth. Acquisition times for one sample can be from 2 to 4 h. This problem is even more acute for in vivo EPR measurements in L-band. Patients cannot be expected to remain stationary for these lengths of time. In order to overcome this limitation, we investigated the dependence of EPR dose measurements on the number of data points in an EPR spectrum. We have shown that this number could be reduced from 1024 to 256 (factor of 4 reduction in spectral acquisition time) at 5 mT magnetic field sweep without a loss of precision in the dose measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.