Abstract

A flow injection system with online sample preparation is proposed for the determination of phosphite in liquid fertilizers by spectrophotometry. After loop-based injection, phosphite is oxidized by an acidic permanganate solution (1.0 10−2 mol L−1 KMnO4 + 1.0 mol L−1 H2SO4) in a heated reactor (50 °C). The phosphate generated is then determined by the molybdenum blue method. Influence of flow rates, temperature, and concentration and order of addition of reagents, sample volume, and reactor configuration for the blue complex formation on recorded signals were investigated. The flow system was applied to phosphite determination in commercial samples of liquid fertilizers. The proposed system handles about 80 samples per hour [0.05–0.40% (w/v) H3PO3; R = 0,9998), consuming about 80 μL sample, 1 mg KMnO4, 25 mg (NH4)6Mo7O24, and 10 mg ascorbic acid per determination. Results are precise [relative standard deviation ≤ 3.5% for 0.1% (w/v) H3PO3, n = 12] and in agreement with those obtained by gravimetry at 95% confidence level. © 2000 John Wiley & Sons, Inc. Lab Robotics and Automation 12:286–290, 2000

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.