Abstract
A spectral resolution measurement of the x-ray imaging crystal spectrometer for the Korea Superconducting Tokamak Advanced Research machine utilizing a segmented position-sensitive detector and time-to-digital converter based delay-line readout electronics was performed by using an x-ray tube in a laboratory. The measured spectral resolution is about 12,600, which means the actual energy resolution is 0.32 eV for the x-ray tube's bremsstrahlung peak energy of 4 keV. The results from the spectral resolution measurement are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.