Abstract
For implant rods as spinal fixation system, low Young's modulus, easy bending to an intended shape (low spring-back), and retaining initial bended shape during a long implantation (low aging-back) are essential properties. The rods were fabricated using a biomedical β-type titanium alloy, Ti–29Nb–13Ta–4.6Zr alloy (TNTZ), and their above-mentioned properties were evaluated in comparison with conventional biomaterials, commercially pure titanium, Ti–6Al–4V ELI, and SUS316L stainless steel. In the results of an animal experiment using implant rods made by TNTZ subjected to solution treatment with low Young's modulus, no negative effects are seen in the radiographs, and no inflammatory cells are observed in histological observation. In spite of low Young's modulus, the spring-back of TNTZ subjected to solution treatment is comparable to that of Ti–6Al–4V ELI. Further, its aging-back is quite small among the other materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.