Abstract

The present work highlights the effect that solvent chemistry has on analytes being detected when utilizing electrospray ionization in ion mobility mass spectrometers. Various studies were conducted involving both positive and negative mode ionization. The enhancements observed by utilizing different levels of acid modifiers as well as the different ionic species of analytes being formed and detected are presented. The makeup of the electrospray solvent determined whether the ionic species would be protonated, sodium adducted, ammonium adducted, proton abstracted or chlorinated. Addition of solvent modifiers such as acids in the positive mode, and organic halides in the negative mode improved signal strength of analytes as well as improved spray characteristics that also decreased signal variability. The ability to perform simultaneous ionization of complex mixtures was also demonstrated through the use of solvent modifiers thereby allowing electrospray ionization ion mobility mass spectrometry the ability to simultaneously analyze complicated mixtures very quickly (millisecond separations).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.