Abstract

Direct, spatially resolved, spectroscopic measurements of C2 in side-heated (spatially isothermal) and Massmann-type (spatially non-isothermal) atomizers showed that the partial pressure of gaseous carbon (pC2) is in thermodynamic, heterogeneous equilibrium for temperatures above 2970 K. No evidence for over-equilibrium excesses of C2 was obtained. The pCN exhibits a gradient, decreasing from the bottom to the top of the graphite tube, indicating the importance of air ingress through the injection hole. Further, the pCN is reduced by the presence of aqueous solutions, primarily close to the point of sample deposition, revealing the non-equilibrium distribution of CN in the atomizer, whereas C2 is only marginally affected. Spatially resolved measurements of Al also indicate the importance of O2 ingress and the sheath gas flow rate on the Al atomic absorption signals, indicating that the pO2 largely determines the degree of Al atom formation and not the presence of gaseous, carbon-containing species.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.