Abstract

Aerosol droplet sizes and velocities from a direct injection nebulizer (DIN) are measured with radial and axial spatial resolution by phase Doppler particle analysis (PDPA). The droplets on the central axis of the spray become finer and their size becomes more uniform when ≍ 20% methanol is added to the usual aqueous solvent. This could explain why the analyte signal is a maximum at this solvent composition when the DIN is used for inductively coupled plasma-mass spectrometry (ICP-MS). Mean droplet velocities are 12 to 22 ms−1 with standard deviations of ±4 to ±7 ms−1. The outer fringes of the aerosol plume tend to be enriched in large droplets. The Sauter mean diameter ( D3,2) and velocity of the droplets also vary substantially with axial position in the aerosol plume.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.