Abstract

The instrument function of the high resolution Thomson scattering (HRTS) diagnostic in the Joint European Torus (JET) has been calculated for use in improved pedestal profile analysis. The full width at half maximum (FWHM) of the spatial instrument response is (22 ± 1) mm for the original HRTS system configuration and depends on the particular magnetic topology of the JET plasmas. An improvement to the optical design of the laser input system is presented. The spatial smearing across magnetic flux surfaces is reduced in this design. The new input system has been implemented (from JPN 78742, July 2009) and the HRTS instrument function corresponding to the new configuration has been improved to approximately FWHM = (9.8 ± 0.8) mm. The reconstructed instrument kernels are used in combination with an ad hoc forward deconvolution procedure for pedestal analysis. This procedure produces good results for both the old and new setups, but the reliability of the deconvolved profiles is greatly reduced when the pedestal width is of the same order as, or less than the FWHM of the instrument kernel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.