Abstract

CCD-based X-ray detector systems often suffer from spatial distortions. Reproducible distortions can be quantitatively measured with a calibration grid and described as a displacement matrix or as spline functions. The measured distortion can be used afterwards to undistort raw images or to refine the actual position of each pixel, e.g. for azimuthal integration. This article describes a method using a regular grid, not necessarily orthogonal, to measure the distortions. The graphical user interface (GUI) Python software that is used to implement this method is available under a GPLv3 license on ESRF GitLab, and produces a spline file that is usable with data-reduction software such as FIT2D or pyFAI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.