Abstract

Sorption of ethylene oxide during and after gaseous sterilization is influenced by numerous factors. It was found that ethylene oxide desorption not only depends on material to be fumigated but also to a considerable degree on the wrapping material. Although polyethylene, polyamide (nylon), polytetrafluoroethylene (Teflon), silicone, aluminum, and glass beads contained no quantities of ethylene oxide detectable by gas chromatography after 72 h of aeration, residual amounts were definitely determined, even after 76 h of aeration in polypropylene, polystyrene, polyvinylchloride, paper products, and compound products of various plastics and paper mixtures. Desorption was, in all cases, found to be better when a mixture of ethylene oxide and methyl formate was used instead of pure ethylene oxide.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.