Abstract

In order to evaluate the sensitivity of ultrasound to intraperitoneal fluid, such as ascites or blood, an experimental study was performed in the pig. Various amounts of fluid were injected into the peritoneal cavity to investigate distribution and diagnostic criteria in different positions. As little as 10 ml of fluid was visualized around the urinary bladder in an upright position. In the supine position, 20 ml could be detected around the bladder and 30 ml around the liver. The injection of at least 60 ml resulted in a pattern of free-floating bowel loops. The sonographic findings of fluid distribution were correlated to radiological studies. As different amounts of fluid produce characteristic sonographic patterns, an approximate estimation of the intraperitoneal fluid volume can be made.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.