Abstract
Some factors affecting the shape of analytical curves in atomic absorption spectroscopy are considered and the influence of the emission and absorption line profiles is discussed in detail. An empirical equation expressing the analytical curves for different ratios of emission line width to absorption line width is given. The possible influence of resonance line broadening and resonance line shift in atomic absorption flame photometry is also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.