Abstract

A rapid, sensitive and rugged solid-phase extraction ultra performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method was developed for determination of paroxetine in human plasma. The procedure for sample preparation includes simple SPE extraction procedure coupled with Hypersil Gold C(18) column (100 mm - 2.1 mm, i.d., 1.9 microm) with isocratic elution at a flow-rate of 0.350 mL/min and fluoxetine was used as the internal standard. The analysis was performed on a triple-quadrupole tandem mass spectrometer by multiple reactions monitoring mode via electrospray ionization. Using 500 microL plasma, the methods were validated over the concentration range 0.050-16.710 ng/mL for paroxetine, with a lower limit of quantification of 0.050 ng/mL. The intra- and inter-day precision and accuracy of the quality control samples were within 10.0%. The recovery was 69.2 and 74.4% for paroxetine and fluoxetine respectively. Total run time was only 1.9 min. The method was highly reproducible and gave peaks with excellent chromatography properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.