Abstract

This paper describes a software correction of scatter coincidence in positron CT which is based on an estimation of scatter projections from true projections by an integral transform. Kernels for the integral transform are projected distributions of scatter coincidences for a line source at different positions in a water phantom and are calculated by Klein-Nishina's formula. True projections of any composite object can be determined from measured projections by iterative applications of the integral transform. The correction method was tested in computer simulations and phantom experiments with Positologica. The results showed that effects of scatter coincidence are not negligible in the quantitation of images, but the correction reduces them significantly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.