Abstract

To increase the sensitivity of measurements of maximal inspiratory pressure (MIP) as a test of inspiratory muscle strength and endurance, software was developed to correct for variation in lung volume. Using a body plethysmograph to determine absolute lung volume during each manoeuvre, values for MIP were expressed as a percentage of the pressure "expected" from the unfatigued MIP/lung volume relation in each subject. The method reduced the variance in peak and average pressure during a series of 18 MIP manoeuvres of 10 seconds duration separated by rest intervals of 10 seconds. The correlation between average pressure and contraction number was improved significantly by the MIP/lung volume correction. This simple correction has many applications in measurements of the MIP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.