Abstract

The accuracy of the new SNAP index with the Bispectral index (BIS) to distinguish different states of propofol/remifentanil anaesthesia was compared in 19 female patients who were undergoing minor gynaecological surgery. Comparisons of the SNAP index, BIS, spectral edge frequency, mean arterial blood pressure and heart rate were performed. The ability of all parameters to distinguish between the steps of anaesthesia -awake vs. loss of response, awake vs. anaesthesia, anaesthesia vs. first reaction and anaesthesia vs. extubation - were analysed with the prediction probability. The prediction probability to differentiate between two interesting nuances of anaesthetic states -loss of response vs. first reaction - was calculated. Only the BIS showed no overlap between the investigated steps of anaesthesia. Both the SNAP index and BIS failed to differentiate the nuances of anaesthesia. The SNAP index and BIS were superior to mean arterial blood pressure and heart rate and spectral edge frequency in distinguishing between different steps of anaesthesia with propofol and remifentanil and provided useful additional information.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.