Abstract

Smart nitinol guidewires have been proposed to improve trackability, facilitating the advancement of catheters through complex vascular anatomies during neurovascular interventions. A smart 0.015 in. diameter nitinol guidewire was actualized through Joule heating of one-way and two-way shape memory alloys (SMA). The device functionalities in terms of bending performance were analyzed: (1) trackability of a 4 Fr catheter as determined in an anatomically correct in vitro environment; (2) time and spatial response of the smart guidewire as a function of material temperature and applied current; and (3) thrombogenic effects as a function of temperature and applied voltage. The results suggest that smart guidewires have substantially improved trackability (i.e., deflection of 15 deg) to overcome the “ledge effect” with the absence of thrombogenicity via a smart guidewire–catheter combined transcatheter based procedure which keeps the catheter surface temperature at 30–33 °C.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.