Abstract

ABSTRACTObjectives: The objective of the study is to evaluate the use of an experimental smart diaper as an indicator of saturation for diaper change in persons with dementia living in nursing homes.Methods: A multicenter prospective study was conducted in 3 nursing homes amongst 18 residents with dementia. For each resident, a frequency–volume urine chart (FVUC) was kept for 24 h including voided volume and diaper weights, wearing smart diapers. A comparative study was set up between results obtained by smart diapers and data registered in FVUCs.Results: Analysis based on quantification of the agreement between saturation calculated by smart diaper and determined by FVUC indicates that measurements reported by sensor do not correspond with measurements based on FVUC. For the regular diaper, the saturation measured by sensor may be 26% below or 39% above saturation based on FVUC and for the super diaper, respectively, 34% below or 30% above.Discussion: This study indicates that the sensor detects and notifies wetness but is not sensitive enough for using it as an indicator for diaper change in people with severe dementia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.