Abstract
In this study, a novel preconcentration/separation technique based on the slurry analysis of chromium loaded on mercaptoundecanoic acid modified TiO2 core-Au shell nanoparticles prior to its determination by electrothermal atomic absorption spectrometry was described. For this purpose, at first, TiO2 nanoparticles were coated with gold shell and then modified with mercaptoundecanoic acid (MUA). Cr (III) was collected on the prepared sorbent by conventional batch technique. After separation of liquid phase, slurry of the sorbent was prepared and directly introduced into graphite furnace of atomic absorption spectrometry. By this way, all drawbacks due to elution procedure were eliminated. Optimum conditions for quantitative sorption and preparation of the slurry were investigated. The chromium in certificated sea-water and spiked drinking water was recovered in the range of 95% confidence level. The proposed technique was fast and simple as well as the risks of contamination and loss during elution were low. The limit of detection (3σ, N=10) was 0.34μgL−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.