Abstract

Twenty-two island skin flaps based on the deep circumflex iliac vessels were elevated in pigs. A comparative study of Laser Doppler flowmetry (LDF) and electromagnetic flowmetry (EMF) was undertaken. Flow values were compared during selective arterial and venous occlusion, nerve stimulation and intra-arterial noradrenaline or papaverine injection. The effectiveness of flap temperatures and blood gas data in monitoring vascular occlusions was also investigated. Similar and simultaneous responses in LDF and EMF values were noted during the different experimental manipulations, but for single simultaneous LDF/EMF observations from different animals the correlation coefficients were poor. The accuracy of EMF measurements in monitoring total flap inflow was confirmed by a high correlation with total venous outflow values. Major vascular occlusions also resulted in significant changes in the flap temperatures, and in the flap blood gases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.