Abstract

Development of GRASE (gradient- and spin-echo) pulse sequences for single-shot 3D imaging has been motivated by physiologic studies of the brain. The duration of echo-planar imaging (EPI) subsequences between RF refocusing pulses in the GRASE sequence is determinant of image distortions and susceptibility artifacts. To reduce these artifacts the regular Cartesian trajectory is modified to a circular trajectory in 2D and a cylindrical trajectory in 3D for reduced echo train time. Incorporation of "fly-back" trajectories lengthened the time of the subsequences and proportionally increased susceptibility artifact but the unipolar readout gradients eliminate all ghost artifacts. The modified cylindrical trajectory reduced susceptibility artifact and distortion artifact while raising the signal-to-noise ratio in both phantom and human brain images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.