Abstract

Rats are used as animal models for many human diseases. Cytokines can serve as biomarkers indicative of these diseases or disease states. Techniques for measuring cytokine expression levels often do not provide the sensitivity needed to measure these biomarkers in biological fluids because the concentrations of many cytokines are below the detection limits of conventional methods. In this paper, we present ultra-sensitive digital immunoassays using Single Molecule Arrays (Simoa) for seven rat cytokines: TNF-α, IL-10, IL-17F, GM-CSF, IFN-γ, IL-4, and IL-1α. These ultra-sensitive immunoassays have limits of detection (LODs) in the femtomolar range and provide the ability to measure rat cytokines in serum below the LODs of conventional immunoassays. We also measured these cytokines in healthy rat serum to obtain baseline levels. The ability to measure cytokines present at low concentrations in rat serum will facilitate future studies of disease using rats as animal models.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.